Ontology highlight
ABSTRACT: Purpose
We report a newly developed device to fragment thrombi in retinal vein occlusion.Methods
The new instrument consists of a 23-gauge (G) pipe and a 37-G needle with an internal wire. A total of 40 porcine eyes were used; 20 eyes for experiments in the branch retinal vein (BRV group) and 20 eyes for experiments in the central retinal vein (CRV group). We placed 25-G 3-port trocars, and core vitrectomy was performed. Another 23-G scleral incision was performed for insertion of the needle. The needle pierced the retinal vein at a distance of three- to four- or one-disc diameters from the optic disc (BRV or CRV group, respectively), and the internal wire was advanced toward the disc. The success rates of needle piercing and cannulation of the internal wire were recorded i
SUBMITTER: Asami T
PROVIDER: S-EPMC5054762 | biostudies-literature | 2016 Sep
REPOSITORIES: biostudies-literature