ABSTRACT: Purpose
Megavoltage computed tomographic (MVCT) imaging has been widely used for the 3-dimensional (3-D) setup of patients treated with helical tomotherapy (HT). One drawback of MVCT is its very long imaging time, the result of slow couch speeds of approximately 1 mm/s, which can be difficult for the patient to tolerate. We sought to develop an MVCT imaging method allowing faster couch speeds and to assess its accuracy for image guidance for HT.Methods and materials
Three cadavers were scanned 4 times with couch speeds of 1, 2, 3, and 4 mm/s. The resulting MVCT images were reconstructed using an iterative reconstruction (IR) algorithm with a penalty term of total variation and with a conventional filtered back projection (FBP) algorithm. The MVCT images were registered with
SUBMITTER: Magome T
PROVIDER: S-EPMC5081222 | biostudies-literature | 2016 Nov
REPOSITORIES: biostudies-literature