Comparison of Polyethylene Wear before and after Hip Revision with Liner Exchange Fixed with the Original Locking Mechanism.


ABSTRACT:

Objective

To compare the wear of conventional ultra-high molecular weight polyethylene (CUHMWPE) and highly cross-linked polyethylene (HCLPE) in hip revision with liner exchange fixed with original locking mechanism using analysis of history medical data.

Methods

From Jan. 1, 2000, to Dec. 31, 2007, 26 patients (with 29 involved hips) underwent liner exchange revision fixed with the original locking mechanism due to wear of CUHMWPE and/or osteolysis. The mean age was 53 ± 9 years at the time of the primary total hip arthroplasty (THA) and 64 ± 9 years at the revision. The exchanged liners (Marathon, Depuy) were made of HCLPE. Annual X-rays were used to measure linear wear and osteolysis. The annual linear penetration was measured using PolyWare® software (Draftware Inc.). An