Ex vivo loading of trussed implants for spine fusion induces heterogeneous strains consistent with homeostatic bone mechanobiology.
Ontology highlight
ABSTRACT: A truss structure was recently introduced as an interbody fusion cage. As a truss system, some of the connected elements may be in a state of compression and others in tension. This study aimed to quantify both the mean and variance of strut strains in such an implant when loaded in a simulated fusion condition with vertebral body or contoured plastic loading platens ex vivo. Cages were each instrumented with 78 fiducial spheres, loaded between platens (vertebral body or contoured plastic), imaged using high resolution micro-CT, and analyzed for deformation and strain of each of the 221 struts. With repeated loading of a cage by vertebral platens, the distribution (variance, indicated by SD) of strut strains widened from 50N control (4±114με, mean±SD) to 1000N (-23±273με) and 2000N (-48±41
SUBMITTER: Caffrey JP
PROVIDER: S-EPMC5164966 | biostudies-literature | 2016 Dec
REPOSITORIES: biostudies-literature
ACCESS DATA