ABSTRACT: Purpose
To compare the effectiveness of prospective, retrospective, and combined (prospective + retrospective) EPI distortion correction methods in bilateral breast diffusion-weighted imaging (DWI) scans.Methods
Five healthy female subjects underwent an axial bilateral breast DWI exam with and without prospective B0 inhomogeneity correction using slice-by-slice linear shimming. In each case, an additional b=0 DWI scan was performed with the polarity of the phase-encoding gradient reversed, to generate an estimated B0 map; this map or a separately acquired B0 map was used for retrospective correction, either alone or in combination with the prospective correction. The alignment between an undistorted, anatomical reference scan with similar co