A Comparison of the Corrected Intraocular Pressure Obtained by the Corvis ST and Reichert 7CR Tonometers in Glaucoma Patients.


ABSTRACT: The purpose of the study was to investigate the accuracy of two corrected intraocular pressure (IOP) measurements by Corvis Scheimpflug Technology (CST)-IOPpachy and by corneal-compensated IOP (IOPcc) using the Reichert 7CR (7CR) tonometers. We also investigated the effects of corneal anatomical and structural parameters on the IOP measurements. The participants included 90 primary open-angle glaucoma patients. We assessed the IOP measurements, obtained by the CST, 7CR, and Goldmann applanation tonometer (GAT), using a paired t-test with Bonferroni correction, Bland-Altman plots, and multiple regression analyses. The 7CR-IOPcc gave the highest value (15.5 ± 2.7 mmHg), followed by the 7CR-IOPg (13.7 ± 3.1 mmHg), GAT-IOP (13.6 ± 2.2 mmHg), CST-IOP (10.3 ± 2.6 mmHg), and CST-IOPpachy (9.7 ± 2