Characterization of Femoral Head Taper Corrosion Features Using a 22-Year Retrieval Database.


ABSTRACT:

Background

Modularity in total hip arthroplasty has been used for decades with great success, but new findings regarding corrosion artifacts have caused a resurgence in tapered junction research. Mechanically assisted crevice corrosion (MACC) is thought to be the mechanism by which corrosive attack occurs. Myriad multi-factorial variables are known to influence the susceptibility of a modular taper junction to MACC. Some of these variables are design and manufacture related and others can be controlled by the surgeon.

Questions/purposes

This study was performed to assess a 22-year retrieval database to determine if correlations exist between severity of corrosion artifacts and head size, time in vivo, head offset, or head material. Secondarily, the agreement of visual and se