Ontology highlight
ABSTRACT: Background
Recent case reports have indicated significant cuff overinflation when using the standard filling volume based on the manufacturer's recommendations in older models of laryngeal tubes. The aim of this study was to determine the minimum cuff pressure needed to perform standardized ventilation without leakage in the new, revised model of the laryngeal tube "LTS-D".Methods
After ethical approval, LTS-D was placed for ventilation in 60 anesthetized patients. The cuff was inflated to the recommended volume (#3: 60 ml, #4: 80 ml, and #5: 90 ml). After evaluation of the initial cuff pressure (CP), the CP was lowered in 10 cmH2O steps until a minimal cuff pressure of 30 cmH2O was achieved. The absence of an audible leak was required for a step-by-st
SUBMITTER: Kriege M
PROVIDER: S-EPMC5290637 | biostudies-literature | 2017 Feb
REPOSITORIES: biostudies-literature