ABSTRACT: Objective
The objective of this study was to determine the safety, feasibility, and efficacy of a predictive hyperglycemia and hypoglycemia minimization (PHHM) system compared with predictive low-glucose insulin suspension (PLGS) alone in overnight glucose control.Research design and methods
A 42-night trial was conducted in 30 individuals with type 1 diabetes in the age range 15-45 years. Participants were randomly assigned each night to either PHHM or PLGS and were blinded to the assignment. The system suspended the insulin pump on both the PHHM and PLGS nights for predicted hypoglycemia but delivered correction boluses for predicted hyperglycemia on PHHM nights only. The primary outcome was the percentage of time spent in a sensor glucose range of 70-180 mg/dL during the
SUBMITTER: Spaic T
PROVIDER: S-EPMC5319476 | biostudies-literature | 2017 Mar