Reproducibility of high-definition (3D) manometry and its agreement with high-resolution (2D) manometry in women with fecal incontinence.


ABSTRACT: While widely used in clinical practice, the reproducibility of high-definition anorectal manometry (HD-ARM) remains unclear. We evaluated the intra-individual reproducibility of HD-ARM and compared pressures measured with HD-ARM and high-resolution anorectal manometry (HR-ARM). Thirty-six women with fecal incontinence had an initial HD-ARM (HD-ARM1); on the same day, after randomization, 21 had a second (HD-ARM2). Sixteen women had a third (HD-ARM3) 4 weeks later. Twenty-five had also been evaluated by HR-ARM previously. Rectoanal pressures were assessed at rest, during squeeze and simulated evacuation. Concordance among pressures was assessed with Lin's concordance correlation coefficient (CCC). Anal resting and squeeze pressures measured with HD-ARM were reproducible on the same and diff