Ontology highlight
ABSTRACT: Background
In Sweden, FreeStyle Libre a flash glucose monitoring system came onto the market in 2014 as a complement to self-monitoring of blood glucose. The aim of this study was to evaluate the accuracy and treatment experience of the FreeStyle Libre system.Methods
Fifty-eight adults with type 1 diabetes used FreeStyle Libre for 10-14 days and measured capillary blood glucose levels with the HemoCue blood glucose measurement system at least six times a day simultaneously.Results
For the entire study period, the mean absolute relative difference (MARD) was 13.2% (95% confidence interval [CI] 12.0%-14.4%). MARD was 13.6% (95% CI 12.1%-15.4%) during week 1 and 12.7% (95% CI 11.5%-13.9%) during week 2. The mean absolute difference (MAD) for the whole study period was
SUBMITTER: Olafsdottir AF