Comparison of self-measured diurnal intraocular pressure profiles using rebound tonometry between primary angle closure glaucoma and primary open angle glaucoma patients.


ABSTRACT:

Purpose

To document the diurnal intraocular pressure (IOP) profile with rebound tonometry performed by primary glaucoma patients in non-clinic environment.

Patients and methods

Fifty-three medically-treated eyes of 31 primary angle closure glaucoma (PACG) and 22 primary open angle glaucoma (POAG) patients with no previous eye surgery were recruited. Diurnal IOP was measured 5 times per day at four-hourly intervals from 08:00 to 24:00 for 1 week in patients' study eye using rebound tonometry in a non-clinic environment. The diurnal IOP profiles were compared between PACG and POAG eyes.

Results

For both PACG and POAG eyes, mean patient-measured IOP was highest in the morning, gradually decreased over the course of a day, and was lowest by midnight (p < 0.001). The diurnal IOP fluctuation ± 1 standard deviation (SD), as documented by SD in daily IOP values, was lower in PACG group (1.6 ± 1.1 mmHg) than in POAG group (2.0 ± 1.2 mmHg; p = 0.049). The mean trough IOP ± 1 SD was higher in PACG group (12.9 ± 2.8 mmHg), compared to POAG group (11.5 ± 3.8 mmHg; p = 0.041). The mean IOP level at midnight ± 1 SD in PACG group (14.0 ± 3.2 mmHg) was higher than that in POAG group (12.1 ± 3.7 mmHg; p = 0.013).

Conclusions

IOP in primary glaucoma patients was highest in the morning, and decreased over the course of a day in non-clinic environment. Treated diurnal IOP fluctuation seemed to be greater in POAG than PACG eyes.