Unknown

Dataset Information

Wash-in and wash-out of sevoflurane in a test-lung model: A comparison between Aisys and FLOW-i.


ABSTRACT: Background: Modern anaesthesia workstations are reassuringly tight and are equipped with effective gas monitoring, thus providing good opportunities for low/minimal flow anaesthesia. A prerequisite for effective low flow anaesthesia is the possibility to rapidly increase and decrease gas concentrations in the circle system, thereby controlling the depth of anaesthesia.  Methods: We studied the wash-in and wash-out of sevoflurane in the circle system with fixed fresh gas flow and vaporizer setting. We compared two modern anaesthesia work stations, the Aisys (GE, Madison, WI, USA) and FLOW-i (Maquet, Solna, Sweden) in a test lung model.  Results: We found fresh-gas flow to have, as expected, a major influence on wash-in, as well as wash-out of sevoflurane. The wash-in ti

SUBMITTER: Jakobsson P 

PROVIDER: S-EPMC5414818 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature