Accuracy Evaluation of a Stereolithographic Surgical Template for Dental Implant Insertion Using 3D Superimposition Protocol.


ABSTRACT:

The aim

of this study was to evaluate the accuracy of a stereolithographic template, with sleeve structure incorporated into the design, for computer-guided dental implant insertion in partially edentulous patients.

Materials and methods

Sixty-five implants were placed in twenty-five consecutive patients with a stereolithographic surgical template. After surgery, digital impression was taken and 3D inaccuracy of implants position at entry point, apex, and angle deviation was measured using an inspection tool software. Mann-Whitney U test was used to compare accuracy between maxillary and mandibular surgical guides. A p value < .05 was considered significant.

Results

Mean (and standard deviation) of 3D error at the entry point was 0.798 mm (±0.52), at the implant apex it was 1.17 mm (±0.63), and mean angular deviation was 2.34 (±0.85). A statistically significant reduced 3D error was observed at entry point p = .037, at implant apex p = .008, and also in angular deviation p = .030 in mandible when comparing to maxilla.

Conclusions

The surgical template used has proved high accuracy for implant insertion. Within the limitations of the present study, the protocol for comparing a digital file (treatment plan) with postinsertion digital impression may be considered a useful procedure for assessing surgical template accuracy, avoiding radiation exposure, during postoperative CBCT scanning.

SUBMITTER: Cristache CM 

PROVIDER: S-EPMC5438863 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy Evaluation of a Stereolithographic Surgical Template for Dental Implant Insertion Using 3D Superimposition Protocol.

Cristache Corina Marilena CM   Gurbanescu Silviu S  

International journal of dentistry 20170507


<h4>The aim</h4>of this study was to evaluate the accuracy of a stereolithographic template, with sleeve structure incorporated into the design, for computer-guided dental implant insertion in partially edentulous patients.<h4>Materials and methods</h4>Sixty-five implants were placed in twenty-five consecutive patients with a stereolithographic surgical template. After surgery, digital impression was taken and 3D inaccuracy of implants position at entry point, apex, and angle deviation was measu  ...[more]

Similar Datasets

| S-EPMC7141331 | biostudies-literature
| S-EPMC6642138 | biostudies-literature
| S-EPMC8852264 | biostudies-literature
| S-EPMC7565125 | biostudies-literature
| S-EPMC5517608 | biostudies-literature
| S-EPMC7142455 | biostudies-literature
| S-EPMC6631019 | biostudies-literature
| S-EPMC6344598 | biostudies-literature
| S-EPMC10439617 | biostudies-literature
| S-EPMC5690229 | biostudies-literature