Unknown

Dataset Information

The unsuitability of implantable Doppler probes for the early detection of renal vascular complications - a porcine model for prevention of renal transplant loss.


ABSTRACT:

Background

Vascular occlusion is a rare, but serious complication after kidney transplantation often resulting in graft loss. We therefore aimed to develop an experimental porcine model for stepwise reduction of the renal venous blood flow and to compare an implantable Doppler probe and microdialysis for fast detection of vascular occlusion.

Methods

In 20 pigs, implantable Doppler probes were placed on the renal artery and vein and a microdialysis catheter was placed in the renal cortex. An arterial flowprobe served as gold standard. Following two-hour baseline measurements, the pigs were randomised to stepwise venous occlusion, complete venous occlusion, complete arterial occlusion or controls.

Results

All parameters were stable through baseline measurements. Glutama

SUBMITTER: Amdisen C 

PROVIDER: S-EPMC5444816 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature