Unknown

Dataset Information

Ambu AuraOnce versus i-gel laryngeal mask airway in infants and children undergoing surgical procedures. A randomized controlled trial.


ABSTRACT:

Objectives

To compare the efficacy and performance of the pediatric Ambu AuraOnce (Ambu AO) mask (Ambu, Copenhagen, Denmark) and i-gel mask (Intersurgical Ltd., Wokingham, United Kingdom). Methods: From May 2015 to September 2016, 112 patients, 0-14 years old, underwent elective surgery at a tertiary university hospital (Riyadh, Saudi Arabia). They were randomly assigned to the Ambu AO or i-gel group. Three groups underwent a subgroup analysis: ≤5 kg (group 1), 5.1-10.0 kg (group 2), and >10 kg (group 3). Results: The oropharyngeal leak pressure was significantly higher for the i-gel (25.4±4.1 cm H2O) than for the Ambu AO (22.5±3.9 cm H2O, p less than 0.001). The Ambu AO had a slightly higher ease of insertion compared to the i-gel (100% versus 94%, p=0.08)  and required less ma

SUBMITTER: Alzahem AM 

PROVIDER: S-EPMC5447208 | biostudies-literature | 2017 May

REPOSITORIES: biostudies-literature