ABSTRACT: Purpose
To improve velocity-to-noise ratio (VNR) and dynamic velocity range of 4D flow magnetic resonance imaging (MRI) by using dual-velocity encoding (dual-venc) with k-t generalized autocalibrating partially parallel acquisition (GRAPPA) acceleration.Materials and methods
A dual-venc 4D flow MRI sequence with k-t GRAPPA acceleration was developed using a shared reference scan followed by three-directional low- and high-venc scans (repetition time / echo time / flip angle = 6.1 msec / 3.4 msec / 15°, temporal/spatial resolution = 43.0 msec/1.2 × 1.2 × 1.2 mm3 ). The high-venc data were used to correct for aliasing in the low-venc data, resulting in a single dataset with the favorable VNR of the low-venc but without velocity aliasing. The sequence was validated