Intentional Laceration of the Anterior Mitral Valve Leaflet to Prevent Left Ventricular Outflow Tract Obstruction During Transcatheter Mitral Valve Replacement: Pre-Clinical Findings.


ABSTRACT:

Objectives

The authors propose a novel transcatheter transection of the anterior mitral leaflet to prevent iatrogenic left ventricular outflow tract (LVOT) obstruction during transcatheter mitral valve replacement (TMVR).

Background

LVOT obstruction is a life-threatening complication of TMVR caused by septal displacement of the anterior mitral leaflet.

Methods

In vivo procedures in swine were guided by biplane x-ray fluoroscopy and intracardiac echocardiography. Retrograde transaortic 6-F guiding catheters straddled the anterior mitral leaflet. A stiff 0.014-inch guidewire with polymer jacket insulation was electrified and advanced from the LVOT, through the A2 leaflet base, into the left atrium. The wire was snared and externalized, forming a loop that was energized

SUBMITTER: Khan JM 

PROVIDER: S-EPMC5476960 | biostudies-literature | 2016 Sep

REPOSITORIES: biostudies-literature