Evaluation of Sensitivity and Specificity Performance of Elecsys HTLV-I/II Assay in a Multicenter Study in Europe and Japan.
Ontology highlight
ABSTRACT: Screening of blood for human T-cell lymphotropic virus type 1 and type 2 (HTLV-1 and -2, respectively) is important to diagnose and prevent infection and ensure the safety of blood supplies. The Elecsys HTLV-I/II assay is a newly developed, electrochemiluminescence screening assay for the detection of HTLV-1/2 infection. The sensitivity and specificity of the Elecsys HTLV-I/II assay were determined using well-characterized HTLV-1/2-positive serum and plasma samples and routine diagnostic and blood donor samples expected to be HTLV negative, respectively. These results were compared with those for at least one of the following CE-marked assays at seven independent laboratories and the Roche Diagnostics facility in Penzberg, Germany: Abbott Architect rHTLV-I/II, Ortho Avioq HTLV-I/II Microel
SUBMITTER: Laperche S