A comparison of interfraction setup error, patient comfort, and therapist acceptance for 2 different prostate radiation therapy immobilization devices.


ABSTRACT:

Purpose

Our purpose was to investigate interfraction setup error of the immobilization device required to implement transperineal ultrasound compared with the current, standard immobilization device. Patient comfort and radiation therapist (RT) satisfaction were also assessed.

Methods and materials

Cone beam computed tomography images were acquired before 4069 fractions from 111 patients (control group, n = 56; intervention group, n = 55) were analyzed. The intervention group was immobilized using the Clarity Immobilization System (CIS), comprising a knee rest with autoscan probe kit and transperineal ultrasound probe (n = 55), and control group using a leg immobilizer (LI) (n = 56). Interfraction setup errors were compared for both groups. Weekly questionnaires using a 10-p