Optical sensing of anticoagulation status: Towards point-of-care coagulation testing.
Ontology highlight
ABSTRACT: Anticoagulant overdose is associated with major bleeding complications. Rapid coagulation sensing may ensure safe and accurate anticoagulant dosing and reduce bleeding risk. Here, we report the novel use of Laser Speckle Rheology (LSR) for measuring anticoagulation and haemodilution status in whole blood. In the LSR approach, blood from 12 patients and 4 swine was placed in disposable cartridges and time-varying intensity fluctuations of laser speckle patterns were measured to quantify the viscoelastic modulus during clotting. Coagulation parameters, mainly clotting time, clot progression rate (α-angle) and maximum clot stiffness (MA) were derived from the clot viscoelasticity trace and compared with standard Thromboelastography (TEG). To demonstrate the capability for anticoagulation sens
SUBMITTER: Tshikudi DM
PROVIDER: S-EPMC5542647 | biostudies-literature | 2017
REPOSITORIES: biostudies-literature