Real-time continuous monitoring of cerebral blood flow autoregulation using near-infrared spectroscopy in patients undergoing cardiopulmonary bypass.


ABSTRACT:

Background and purpose

Individualizing mean arterial blood pressure targets to a patient's cerebral blood flow autoregulatory range might prevent brain ischemia for patients undergoing cardiopulmonary bypass (CPB). This study compares the accuracy of real-time cerebral blood flow autoregulation monitoring using near-infrared spectroscopy with that of transcranial Doppler.

Methods

Sixty adult patients undergoing CPB had transcranial Doppler monitoring of middle cerebral artery blood flow velocity and near-infrared spectroscopy monitoring. The mean velocity index (Mx) was calculated as a moving, linear correlation coefficient between slow waves of middle cerebral artery blood flow velocity and mean arterial blood pressure. The cerebral oximetry index was calculated as a simila