Dependence of Boiling Histotripsy Treatment Efficiency on HIFU Frequency and Focal Pressure Levels.


ABSTRACT: Boiling histotripsy (BH) is a high-intensity focused ultrasound (HIFU)-based method of mechanical tissue fractionation that utilizes millisecond-long bursts of HIFU shock waves to cause boiling at the focus in milliseconds. The subsequent interaction of the incoming shocks with the vapor bubble mechanically lyses surrounding tissue and cells. The acoustic parameter space for BH has been investigated previously and an inverse dependence between the HIFU frequency and the dimensions of a BH lesion has been observed. The primary goal of the present study was to investigate in more detail the ablation rate and reliability of BH in the frequency range relevant to treatment of deep abdominal tissue targets (1-2 MHz). The second goal was to investigate the effect of focal peak pressure levels and