ABSTRACT: This study investigated the applicability of an HY-differential mobility analyzer with an optical particle counter (HY-DMA/OPC), named as KOFAM, for counting fibrous matters in real time. Fibers separated from particles by the HY-DMA were counted with an OPC. To assess the KOFAM performance, the proposed method and the conventional gold standard phase contrast microscopy (PCM) method were compared in terms of variables such as recovery, relative difference, coefficient of determination, and conformity. The optimal sheath-to-aerosol (outlet) flow ratio of the internal flow in the HY-DMA was determined to be 1.6:1. In terms of recovery of the HY-DMA, the highest recovery was obtained at a voltage of 500 V regardless of which type of asbestos was tested. The recovery rate for serpentine was 4