ABSTRACT: Purpose
To improve the precision of proton resonance frequency-shift magnetic resonance thermometry near ablation probes by recovering near-probe image signals that are typically lost due to magnetic susceptibility-induced field distortions.Methods
A dual-echo gradient-recalled echo sequence was implemented, in which the first echo was under- or over-refocused in the slice dimension to recover image signal and temperature precision near a probe, and the second echo was fully refocused to obtain image signal everywhere else in the slice. A penalized maximum likelihood algorithm was implemented to estimate a single temperature map from both echoes. Agar phantom and ex vivo experiments with and without microwave heating at 3 T evaluated how much temperature precision was impro