Improved usability of a multi-infusion setup using a centralized control interface: A task-based usability test.


ABSTRACT: The objective of this study was to assess the usability benefits of adding a bedside central control interface that controls all intravenous (IV) infusion pumps compared to the conventional individual control of multiple infusion pumps. Eighteen dedicated ICU nurses volunteered in a between-subjects task-based usability test. A newly developed central control interface was compared to conventional control of multiple infusion pumps in a simulated ICU setting. Task execution time, clicks, errors and questionnaire responses were evaluated. Overall the central control interface outperformed the conventional control in terms of fewer user actions (40±3 vs. 73±20 clicks, p<0.001) and fewer user errors (1±1 vs. 3±2 errors, p<0.05), with no difference in task execution times (421±108 vs. 406±119