Single-scan rest/stress imaging with 99mTc-Sestamibi and cadmium zinc telluride-based SPECT for hyperemic flow quantification: A feasibility study evaluated with cardiac magnetic resonance imaging.


ABSTRACT:

Introduction

We aimed to evaluate whether the hyperemic myocardial blood flow (MBF) can be estimated using cadmium zinc telluride (CZT)-based single-photon emission computed tomography (SPECT) cameras with a single, rapid rest/stress dynamic scan. Dynamic contrast-enhanced (DCE) cardiac magnetic resonance imaging (MRI) was used as a reference modality for flow measurement.

Materials and methods

The proposed protocol included both the rest and stress acquisitions within a 24-min scan. Patients were first injected with 99mTc-Sestamibi at the resting state. Sixty minutes after the first injection, the subject was positioned via scintigraphy, after which the list-mode data acquisition was initiated and continued for 24 minutes. Five minutes after data acquisition was initiated,