Quantitative Image Quality Comparison of Reduced- and Standard-Dose Dual-Energy Multiphase Chest, Abdomen, and Pelvis CT.
Ontology highlight
ABSTRACT: We present a new image quality assessment method for determining whether reducing radiation dose impairs the image quality of computed tomography (CT) in qualitative and quantitative clinical analyses tasks. In this Institutional Review Board-exempt study, we conducted a review of 50 patients (male, 22; female, 28) who underwent reduced-dose CT scanning on the first follow-up after standard-dose multiphase CT scanning. Scans were for surveillance of von Hippel-Lindau disease (N = 26) and renal cell carcinoma (N = 10). We investigated density, morphometric, and structural differences between scans both at tissue (fat, bone) and organ levels (liver, heart, spleen, lung). To quantify structural variations caused by image quality differences, we propose using the following metrics: dice simila
SUBMITTER: Buty M
PROVIDER: S-EPMC5573232 | biostudies-literature | 2017 Jun