Hypo-Attenuated Leaflet Thickening and Reduced Leaflet Motion in Sutureless Bioprosthetic Aortic Valves.
Ontology highlight
ABSTRACT: The prevalence of hypo-attenuated leaflet thickening (HALT) and reduced leaflet motion (RLM) is unknown in surgically implanted bioprostheses because systematic investigation of HALT and/or RLM is limited to a few catheter-based valves. The aim of this study was to investigate the prevalence of HALT and RLM by cardiac computed tomography in patients who underwent surgical aortic valve replacement and received a Perceval sutureless aortic valve bioprosthesis. This was a single-center prospective observational study that included 47 patients (83.5% of the total number of implantations) who underwent surgical aortic valve replacement with implantation of the Perceval sutureless bioprosthesis (LivaNova PLC, London, UK) at Karolinska University Hospital in Stockholm, Sweden from 2012 to 2016 an