Video Head Impulse Tests with a Remote Camera System: Normative Values of Semicircular Canal Vestibulo-Ocular Reflex Gain in Infants and Children.
Ontology highlight
ABSTRACT: The video head impulse test (VHIT) is widely used to identify semicircular canal function impairments in adults. But classical VHIT testing systems attach goggles tightly to the head, which is not tolerated by infants. Remote video detection of head and eye movements resolves this issue and, here, we report VHIT protocols and normative values for children. Vestibulo-ocular reflex (VOR) gain was measured for all canals of 303 healthy subjects, including 274 children (aged 2.6 months-15 years) and 26 adults (aged 16-67). We used the Synapsys® (Marseilles, France) VHIT Ulmer system whose remote camera measures head and eye movements. HITs were performed at high velocities. Testing typically lasts 5-10 min. In infants as young as 3 months old, VHIT yielded good inter-measure replica
SUBMITTER: Wiener-Vacher SR
PROVIDER: S-EPMC5594068 | biostudies-literature | 2017
REPOSITORIES: biostudies-literature
ACCESS DATA