Unknown

Dataset Information

Correlation of optic nerve sheath diameter with directly measured intracranial pressure in Korean adults using bedside ultrasonography.


ABSTRACT:

Objectives

The correlation of optic nerve sheath diameter (ONSD) as seen on ultrasonography (US) and directly measured intracranial pressure (ICP) has been well described. Nevertheless, differences in ethnicity and type of ICP monitor used are obstacles to the interpretation. Therefore, we investigated the direct correlation between ONSD and ventricular ICP and defined an optimal cut-off point for identifying increased ICP (IICP) in Korean adults with brain lesions.

Methods

This prospective study included patients who required an external ventricular drainage (EVD) catheter for ICP control. IICP was defined as an opening pressure over 20 mmHg. ONSD was measured using a 13 MHz US probe before the procedure. Linear regression analysis and receiver operator characteristic (ROC)

SUBMITTER: Jeon JP 

PROVIDER: S-EPMC5597178 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature