Robustness of two different methods of monitoring respiratory system compliance during mechanical ventilation.


ABSTRACT: Robustness measures the performance of estimation methods when they work under non-ideal conditions. We compared the robustness of artificial neural networks (ANNs) and multilinear fitting (MLF) methods in estimating respiratory system compliance (C RS) during mechanical ventilation (MV). Twenty-four anaesthetized pigs underwent MV. Airway pressure, flow and volume were recorded at fixed intervals after the induction of acute lung injury. After consecutive mechanical breaths, an inspiratory pause (BIP) was applied in order to calculate CRS using the interrupter technique. From the breath preceding the BIP, ANN and MLF had to compute CRS in the presence of two types of perturbations: transient sensor disconnection (TD) and random noise (RN). Performance of t