A comparison of the non-invasive ultrasonic cardiac output monitor (USCOM) with the oesophageal Doppler monitor during major abdominal surgery.


ABSTRACT:

Background

Perioperative interventions, targeted to increase global blood flow defined by explicit measured goals, reduce postoperative complications. Consequently, reliable non-invasive estimation of the cardiac output could have far-reaching benefit.

Methods

This study compared a non-invasive Doppler device - the ultrasonic cardiac output monitor (USCOM) - with the oesophageal Doppler monitor (ODM), on 25 patients during major abdominal surgery. Stroke volume was determined by USCOM (SVUSCOM) and ODM (SVODM) pre and post fluid challenges.

Results

A ≥ 10% change (Δ) SVUSCOM had a sensitivity of 94% and specificity of 88% to detect a ≥ 10% Δ SVODM; the area under the receiver operating curve was 0.94 (95% CI 0.90-0.99). Conc