ABSTRACT: Purpose
To evaluate the agreement of cycloplegic refractive error measures between the Grand Seiko and Retinomax autorefractors in 4- and 5-year-old children.Methods
Cycloplegic refractive error of children was measured using the Grand Seiko and Retinomax during a comprehensive eye examination. Accommodative error was measured using the Grand Seiko. The differences in sphere, cylinder, spherical equivalent (SE) and intereye vector dioptric distance (VDD) between autorefractors were assessed using the Bland-Altman plot and 95% limits of agreement (95% LoA).Results
A total of 702 examinations were included. Compared to the Retinomax, the Grand Seiko provided statistically significantly larger values of sphere (mean difference, 0.34 D; 95% LoA, -0.46 to 1.14 D), SE (me