Ontology highlight
ABSTRACT: Objectives
To benchmark CT-dose data for standard adult CT studies to regional and national reference levels using a dose-tracking system.Methods
Data from five CT systems from three hospitals were collected over a 1- to 2.5-year period (2012-2014), using the same type of dose management system. Inclusion criteria were adult patients and standard CT-head, CT-abdomen-pelvis, CT-thorax, CT-lumbar spine, CT-pulmonary embolism, CT-cervical spine and CT-thorax-abdomen studies, with one helical scan. Volumetric CT-dose index (CTDIvol), dose length product (DLP) and scan length from 31,709 scans were analysed statistically.Results
After dose optimisation CTDIvol and DLP values were below the national diagnostic reference levels (DRLs) for all CT studi
SUBMITTER: Pyfferoen L
PROVIDER: S-EPMC5621994 | biostudies-literature | 2017 Oct
REPOSITORIES: biostudies-literature