Benchmarking adult CT-dose levels to regional and national references using a dose-tracking software: a multicentre experience.


ABSTRACT:

Objectives

To benchmark CT-dose data for standard adult CT studies to regional and national reference levels using a dose-tracking system.

Methods

Data from five CT systems from three hospitals were collected over a 1- to 2.5-year period (2012-2014), using the same type of dose management system. Inclusion criteria were adult patients and standard CT-head, CT-abdomen-pelvis, CT-thorax, CT-lumbar spine, CT-pulmonary embolism, CT-cervical spine and CT-thorax-abdomen studies, with one helical scan. Volumetric CT-dose index (CTDIvol), dose length product (DLP) and scan length from 31,709 scans were analysed statistically.

Results

After dose optimisation CTDIvol and DLP values were below the national diagnostic reference levels (DRLs) for all CT studi