A pilot evaluation of whole blood finger-prick sampling for point-of-care HIV viral load measurement: the UNICORN study.
Ontology highlight
ABSTRACT: There is a global need for HIV viral load point-of-care (PoC) assays to monitor patients receiving antiretroviral therapy. UNICORN was the first study of an off-label protocol using whole blood finger-prick samples tested with and without a simple three minute spin using a clinic-room microcentrifuge. Two PoC assays were evaluated in 40 HIV-positive participants, 20 with detectable and 20 with undetectable plasma viral load (pVL) (<20 copies/ml). Using 100 µl finger-prick blood samples, the Cepheid Xpert HIV-1 Viral Load and HIV-1 Qual cartridges were compared with laboratory pVL assessment (TaqMan, Roche). For participants with undetectable viraemia by TaqMan, there was poor concordance without centrifugation with the TaqMan platform with only 40% 'undetectable' using Xpert VL and 25% 'no