Novel volumetric method for highly repeatable injection in microchip electrophoresis.
Ontology highlight
ABSTRACT: A novel injector for microchip electrophoresis (MCE) has been designed and evaluated that achieves very high repeatability of injection volume suitable for quantitative analysis. It eliminates the injection biases in electrokinetic injection and the dependence on pressure and sample properties in hydrodynamic injection. The microfluidic injector, made of poly(dimethylsiloxane) (PDMS), operates similarly to an HPLC injection valve. It contains a channel segment (chamber) with a well-defined volume that serves as an "injection loop". Using on-chip microvalves, the chamber can be connected to the sample source during the "loading" step, and to the CE separation channel during the "injection" step. Once the valves are opened in the second state, electrophoretic potential is applied to separate
SUBMITTER: Ha NS
PROVIDER: S-EPMC5657552 | biostudies-literature | 2017 Sep
REPOSITORIES: biostudies-literature
ACCESS DATA