Unknown

Dataset Information

Validation of a light-scattering PM2.5 sensor monitor based on the long-term gravimetric measurements in field tests.


ABSTRACT:

Background

Portable direct-reading instruments by light-scattering method are increasingly used in airborne fine particulate matter (PM2.5) monitoring. However, there are limited calibration studies on such instruments by applying the gravimetric method as reference method in field tests.

Methods

An 8-month sampling was performed and 96 pairs of PM2.5 data by both the gravimetric method and the simultaneous light-scattering real-time monitoring (QT-50) were obtained from July, 2015 to February, 2016 in Shanghai. Temperature and relative humidity (RH) were recorded. Mann-Whitney U nonparametric test and Spearman correlation were used to investigate the differences between the two measurements. Multiple linear regression (MLR) model was applied to set up the calibration model

SUBMITTER: Shi J 

PROVIDER: S-EPMC5679553 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature