ABSTRACT: Background
In field studies, hemoglobin (Hb) is often measured using a battery-operated, portable HemoCue® hemoglobinometer.Methods
We compared the performance of 2 HemoCue® models (Hb-201+ and Hb-301) and investigated effects of preanalytical factors on Hb results by simulating unfavorable field conditions.Results
The Hb-301 produced 2.6% higher results compared to the Hb-201+. Hb had to be measured within 1min of filling the Hb-301 cuvette to avoid artificially elevated concentrations (1.3% per min). The Hb-301 cuvettes withstood elevated temperature (37°C) and humidity (72%) for 3weeks, while the Hb-201+ cuvettes degraded within 10min under those conditions. Both cuvette types withstood elevated temperature for 3weeks. Properly-collected venous and capillary bloo