Radiation dose reduction in thoracic and abdomen-pelvic CT using tube current modulation: a phantom study.
Ontology highlight
ABSTRACT: This phantom study was designed to compare the radiation dose in thoracic and abdomen-pelvic CT scans with and without use of tube current modulation (TCM). Effective dose (ED) and size-specific dose estimation (SSDE) were calculated with the absorbed doses measured at selective radiosensitive organs using a thermoluminescence dosimeter-100 (TLD-100). When compared to protocols without TCM, the ED and SSDE were reduced significantly with use of TCM for both the thoracic and abdomen-pelvic CT. With use of TCM, the ED was 6.50 ± 0.29 mSv for thoracic and 6.01 ± 0.20 mSv for the abdomen-pelvic CT protocols. However without use of TCM, the ED was 20.07 ± 0.24 mSv and 17.30 ± 0.41 mSv for the thoracic and abdomen-pelvic CT protocols, respectively. The corresponding SSDE was 10.18 ± 0.48 mGy and