Dosimetric evaluation of the Acuros XB algorithm for a 4 MV photon beam in head and neck intensity-modulated radiation therapy.
Ontology highlight
ABSTRACT: In this study, we assessed the differences in the dose distribution of a 4 MV photon beam among different calculation algorithms: the Acuros XB (AXB) algorithm, the analytic anisotropic algorithm (AAA), and the pencil beam convolution (PBC) algorithm (ver. 11.0.31), in phantoms and in clinical intensity-modulated radiation therapy (IMRT) plans. Homogeneous and heterogeneous, including middle-, low-, and high-density, phantoms were combined to assess the percentage depth dose and lateral dose profiles among AXB, AAA, and PBC. For the phantom containing the low-density area, AXB was in agreement with measurement within 0.5%, while the greatest differences between the AAA and PBC calculations and measurement were 2.7% and 3.6%, respectively. AXB showed agreement with measurement within 2.5% a