Unknown

Dataset Information

Electrocardiogram-gated coronary CT angiography dose estimates using ImPACT.


ABSTRACT: The primary study objective was to assess radiation doses using a modified form of the Imaging Performance Assessment of Computed Tomography (CT) scanner (ImPACT) patient dosimetry for cardiac applications on an Aquilion ONE ViSION Edition scanner, including the Ca score, target computed tomography angiography (CTA), prospective CTA, continuous CTA/cardiac function analysis (CFA), and CTA/CFA modulation. Accordingly, we clarified the CT dose index (CTDI) to determine the relationship between heart rate (HR) and X-ray exposure. As a secondary objective, we compared radiation doses using modified ImPACT, a whole-body dosimetry phantom study, and the k-factor method to verify the validity of the dose results obtained with modified ImPACT. The effective dose determined for the reference person

SUBMITTER: Kobayashi M 

PROVIDER: S-EPMC5690042 | biostudies-literature | 2016 Jul

REPOSITORIES: biostudies-literature