Commissioning a 50-100 kV X-ray unit for skin cancer treatment.


ABSTRACT: This study provides the authors' experience along with dosimetric data from the commissioning of two Sensus SRT-100 50-100 kV X-ray units. Data collected during the commissioning process included: a) HVL, b) output (dose rate), c) applicator cone factors, and d) percentage depth dose. A Farmer-type chamber (PTW-N23333), and a thin-window parallel plate ion chamber (PTW-N23342) were used for dose rate measurements and dose profiles were measured with EBT3 GafChromic film. The average HVL values for 50, 70, and 100 kV of the two treatment units were found to be 0.52, 1.15, and 2.20 mm Al, respectively. The HVL's were 5%-9% lower when measured with the Farmer chamber, as compared to measurements with the parallel plate chamber, for energies of 70 and 100 kV. Dose rates were also measured to b