Initial clinical experience with a radiation oncology dedicated open 1.0T MR-simulation.
Ontology highlight
ABSTRACT: The purpose of this study was to describe our experience with 1.0T MR-SIM including characterization, quality assurance (QA) program, and features necessary for treatment planning. Staffing, safety, and patient screening procedures were developed. Utilization of an external laser positioning system (ELPS) and MR-compatible couchtop were illustrated. Spatial and volumetric analyses were conducted between CT-SIM and MR-SIM using a stereotactic QA phantom with known landmarks and volumes. Magnetic field inhomogeneity was determined using phase difference analysis. System-related, in-plane distortion was evaluated and temporal changes were assessed. 3D distortion was characterized for regions of interest (ROIs) 5-20 cm away from isocenter. American College of Radiology (ACR) recommended tests