Estimation of patient setup uncertainty using BrainLAB Exatrac X-Ray 6D system in image-guided radiotherapy.


ABSTRACT: The purpose of this study was to evaluate setup uncertainties for brain sites with ExacTrac X-Ray 6D system and to provide optimal margin guidelines. Fifteen patients with brain tumor were included in this study. Two X-ray images with ExacTrac X-Ray 6D system were used to verify patient position and tumor target localization before each treatment. The 6D fusion software first generates various sets of DRRs with position variations in both three translational and three rotational directions (six degrees of freedom) for the CT images. Setup variations (translation and rotation) after correction were recorded and corrected before treatment. The 3D deviations are expressed as mean ± standard deviation. The random error (Σ(σi)), systematic error (μi), and group systematic error (M(μi)) for the