Computed tomography dosimetry with high-resolution detectors commonly used in radiotherapy - an energy dependence study.
Ontology highlight
ABSTRACT: New methods of dosimetry in computed tomography (CT) X-ray fields require the use of high-resolution detectors instead of pencil-type ionization chambers typically used for CT dose index (CTDI) measurements. This paper presents a study on the suitability of a wide range of ionization chambers, diodes, and a two-dimensional detector array, used primarily in radiation therapy, for CT and cone-beam CT dosimetry. Specifically, the energy dependence of these detectors from 50 kVp up to 125 kVp is reported. All measurements were performed in reference to a calibrated diode for use in this energy region. The radiation quality correction factors provided by the manufacturer were used, depending on the measured half-value layer (HVL) for the particular X-ray beam. Our study demonstrated the general
SUBMITTER: Liebmann M
PROVIDER: S-EPMC5690150 | biostudies-literature | 2015 Sep
REPOSITORIES: biostudies-literature
ACCESS DATA