Evaluation of MVCT imaging dose levels during helical IGRT: comparison between ion chamber, TLD, and EBT3 films.
Ontology highlight
ABSTRACT: The purpose of this investigation was to evaluate the dose on megavoltage CT (MVCT) images required for tomotherapy. As imaging possibilities are often used before each treatment and usually used several times before the session, we tried to evaluate the dose delivered during the procedure. For each scanning mode (fine, normal, and coarse), we first established the relative variation of these doses according to different technical parameters (explored length, patient setup). These dose variations measured with the TomoPhant, also known as Cheese phantom, showed the expected variations (due to the variation of scattered radiation) of 15% according to the explored length and ± 5% according to the phantom setup (due to the variation of the point of measurement in the bore). In order to estima