Unknown

Dataset Information

Evaluation of commercial extension plates for the ACR CT accreditation phantom.


ABSTRACT: The American College of Radiology (ACR) Computed Tomography (CT) Accreditation Program requires submission of phantom scans acquired with the ACR accreditation phantom. There is a known issue with some wide-beam scanners in which the Hounsfield unit (HU) value of water may be correct when using the scanner manufacturer's phantom, but will be out of range in some scan modes when scanning the accreditation phantom. The phantom manufacturer has developed a product known as Extension Plates to eliminate the water HU value issue. The purpose of this technical note is to evaluate the effectiveness of the Extension Plates in alleviating the water HU issue. The ACR phantom was scanned on nine different CT scanners representing four CT manufacturers at eight different facilities. Scanner models inc

SUBMITTER: Greene-Donnelly KA 

PROVIDER: S-EPMC5690207 | biostudies-literature | 2016 Jan

REPOSITORIES: biostudies-literature