Evaluation of digital radiography practice using exposure index tracking.


ABSTRACT: Some digital radiography (DR) detectors and software allow for remote download of exam statistics, including image reject status, body part, projection, and exposure index (EI). The ability to have automated data collection from multiple DR units is conducive to a quality control (QC) program monitoring institutional radiographic exposures. We have implemented such a QC program with the goal to identify outliers in machine radiation output and opportunities for improvement in radiation dose levels. We studied the QC records of four digital detectors in greater detail on a monthly basis for one year. Although individual patient entrance skin exposure varied, the radiation dose levels to the detectors were made to be consistent via phototimer recalibration. The exposure data stored on each d