Dosimetric properties of equivalent-quality flattening filter-free (FFF) and flattened photon beams of Versa HD linear accelerator.
Ontology highlight
ABSTRACT: This study presents the basic dosimetric properties of photon beams of a Versa HD linear accelerator (linac), which is capable of delivering flattening filter-free (FFF) beams with a beam quality equivalent to the corresponding flattened beams based on comprehensive beam data measurement. The analyzed data included the PDDs, profiles, penumbra, out-of-field doses, surface doses, output factors, head and phantom scatter factors, and MLC transmissions for both FFF and flattened beams of 6 MV and 10 MV energy from an Elekta Versa HD linac. The 6MVFFF and 10MVFFF beams had an equivalent mean energy to the flattened beams and showed less PDD variations with the field sizes. Compared with their corresponding flattened beams, Dmax was deeper for FFF beams for all field sizes; the ionization ratio
SUBMITTER: Meshram MN
PROVIDER: S-EPMC5690903 | biostudies-literature | 2016 May
REPOSITORIES: biostudies-literature